Anthropomorphic lung phantom based validation of in-room proton therapy 4D-CBCT image correction for dose calculation
- Source :
- Zeitschrift fur Medizinische Physik, Zeitschrift fur Medizinische Physik, Elsevier, In press, ⟨10.1016/j.zemedi.2020.09.004⟩, Z. Med. Phys. 32, 74-84 (2022)
- Publication Year :
- 2020
- Publisher :
- HAL CCSD, 2020.
-
Abstract
- Purpose Ventilation-induced tumour motion remains a challenge for the accuracy of proton therapy treatments in lung patients. We investigated the feasibility of using a 4D virtual CT (4D-vCT) approach based on deformable image registration (DIR) and motion-aware 4D CBCT reconstruction (MA-ROOSTER) to enable accurate daily proton dose calculation using a gantry-mounted CBCT scanner tailored to proton therapy. Methods Ventilation correlated data of 10 breathing phases were acquired from a porcine ex-vivo functional lung phantom using CT and CBCT. 4D-vCTs were generated by (1) DIR of the mid-position 4D-CT to the mid-position 4D-CBCT (reconstructed with the MA-ROOSTER) using a diffeomorphic Morphons algorithm and (2) subsequent propagation of the obtained mid-position vCT to the individual 4D-CBCT phases. Proton therapy treatment planning was performed to evaluate dose calculation accuracy of the 4D-vCTs. A robust treatment plan delivering a nominal dose of 60 Gy was generated on the average intensity image of the 4D-CT for an approximated internal target volume (ITV). Dose distributions were then recalculated on individual phases of the 4D-CT and the 4D-vCT based on the optimized plan. Dose accumulation was performed for 4D-vCT and 4D-CT using DIR of each phase to the mid position, which was chosen as reference. Dose based on the 4D-vCT was then evaluated against the dose calculated on 4D-CT both, phase-by-phase as well as accumulated, by comparing dose volume histogram (DVH) values (Dmean, D2%, D98%, D95%) for the ITV, and by a 3D-gamma index analysis (global, 3%/3 mm, 5 Gy, 20 Gy and 30 Gy dose thresholds). Results Good agreement was found between the 4D-CT and 4D-vCT-based ITV-DVH curves. The relative differences ((CT-vCT)/CT) between accumulated values of ITV Dmean, D2%, D95% and D98% for the 4D-CT and 4D-vCT-based dose distributions were −0.2%, 0.0%, −0.1% and −0.1%, respectively. Phase specific values varied between −0.5% and 0.2%, −0.2% and 0.5%, −3.5% and 1.5%, and −5.7% and 2.3%. The relative difference of accumulated Dmean over the lungs was 2.3% and Dmean for the phases varied between −5.4% and 5.8%. The gamma pass-rates with 5 Gy, 20 Gy and 30 Gy thresholds for the accumulated doses were 96.7%, 99.6% and 99.9%, respectively. Phase-by-phase comparison yielded pass-rates between 86% and 97%, 88% and 98%, and 94% and 100%. Conclusions Feasibility of the suggested 4D-vCT workflow using proton therapy specific imaging equipment was shown. Results indicate the potential of the method to be applied for daily 4D proton dose estimation.
- Subjects :
- Male
Dose-volume histogram
Lung Neoplasms
Swine
Biophysics
Image registration
Imaging phantom
030218 nuclear medicine & medical imaging
03 medical and health sciences
4d-vct
Cone-beam
Motion
Proton Therapy
Thorax
Tomography
0302 clinical medicine
Image Processing, Computer-Assisted
[INFO.INFO-IM]Computer Science [cs]/Medical Imaging
Animals
Humans
Medicine
Radiology, Nuclear Medicine and imaging
Four-Dimensional Computed Tomography
Radiation treatment planning
Lung
Proton therapy
ComputingMilieux_MISCELLANEOUS
Radiological and Ultrasound Technology
Phantoms, Imaging
business.industry
Radiotherapy Planning, Computer-Assisted
Radiotherapy Dosage
Spiral Cone-Beam Computed Tomography
Cone-Beam Computed Tomography
Intensity (physics)
medicine.anatomical_structure
business
Nuclear medicine
Chickens
Subjects
Details
- Language :
- English
- ISSN :
- 09393889
- Database :
- OpenAIRE
- Journal :
- Zeitschrift fur Medizinische Physik, Zeitschrift fur Medizinische Physik, Elsevier, In press, ⟨10.1016/j.zemedi.2020.09.004⟩, Z. Med. Phys. 32, 74-84 (2022)
- Accession number :
- edsair.doi.dedup.....7ced982335637f08b25e0c2f2e29811b
- Full Text :
- https://doi.org/10.1016/j.zemedi.2020.09.004⟩