Analytical Performance and Quality Indicators of Fully Automated Immunoassays for sFlt-1 and PlGF
- Source :
- The journal of applied laboratory medicine. 7(2)
- Publication Year :
- 2021
-
Abstract
- Background The ratio of the antiangiogenic factor, soluble fms-like tyrosine kinase 1 (sFlt-1), to the proangiogenic factor, placental growth factor (PlGF), is associated with increased risk of preeclampsia. Here, we describe an analytical evaluation of the Elecsys sFlt-1 and PlGF assays at the first North American site in which they were clinically implemented. Methods The analytical evaluation included short- and long-term imprecision, method comparison, accuracy, linearity, sample stability, limit of quantification verification, and measurement uncertainty. Quality indicators were also evaluated, including turnaround time and repeat test frequency. Results Short-term (13-day) and long-term (12-month) imprecision for sFlt-1 and PlGF were 0.99 and bias 38, >85, and >110) exhibited negative percent agreement and positive percent agreement of 100%, respectively (n = 40). During the first 12 months, 257 orders were placed, repeat test frequency was 17.5%, mean time between repeat orders was 23 days, and 72.0% of results were reported within 2 h from sample receipt when quality control was run continuously. Conclusions We describe analytical performance parameters and quality indicators of the Elecsys sFlt-1 and PlGF assays, which was the first North American clinical laboratory site to implement these assays in support of the institution’s high-risk obstetrical unit.
Details
- ISSN :
- 25769456
- Volume :
- 7
- Issue :
- 2
- Database :
- OpenAIRE
- Journal :
- The journal of applied laboratory medicine
- Accession number :
- edsair.doi.dedup.....7cbea542a442933ee15a541cde38ea55