Accuracy of user-adjusted axial length measurements with optical biometry in eyes having combined phacovitrectomy for macular-off rhegmatogenous retinal detachment
- Source :
- Journal of Cataract and Refractive Surgery. 42:1009-1014
- Publication Year :
- 2016
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 2016.
-
Abstract
- Purpose To evaluate the accuracy of user-adjusted axial length (AL) measured by optical biometry for intraocular lens (IOL) calculations in eyes having combined phacovitrectomy for macula-off rhegmatogenous retinal detachment (RRD). Setting Ophthalmology Department, Calderdale Royal Hospital, Halifax, United Kingdom. Design Retrospective case series. Methods Consecutive eyes having phacovitrectomy for macula-off RRD were evaluated. The ALs were measured using optical biometry with user adjustment to identify a posterior peak corresponding to the eye's AL and ultrasound (US). These were compared with each other and with the postoperative optical biometry and analyzed for accuracy as an indication of the eye's AL. Results The study comprised 22 eyes of 22 patients. There was no statistically significant difference between the mean AL measurements derived from user-adjusted optical biometry and US AL ( P = .964). The user-adjusted optical biometry was not statistically significantly different from the postoperative optical biometry ( P = .242). Compared with the postoperative optical biometry, the IOL power was within ±0.5 diopter in 12 (92%) of 13 cases (95% confidence interval [CI], 77.8 to 100.0) for user-adjusted optical biometry and in 10 (77%) of 13 cases (95% CI, 54.0 to 99.8) for US measurements. Conclusions User-adjusted optical biometry could be used as an alternative method for measuring AL in macula-off RRD with combined phacovitrectomy. However, optical biometry would require assessment of agreement with US AL in cases in which a posterior peak is not easily identifiable. User-adjusted optical biometry might outperform US AL when calculating IOL power; however, a larger study should be performed to confirm this. Financial Disclosure None of the authors has a financial or proprietary interest in any material or method mentioned.
- Subjects :
- 0301 basic medicine
Optics and Photonics
medicine.medical_specialty
Biometry
medicine.medical_treatment
Intraocular lens
Refraction, Ocular
Q1
03 medical and health sciences
0302 clinical medicine
Optical biometry
Lens Implantation, Intraocular
Ophthalmology
medicine
Humans
Dioptre
Retrospective Studies
Lenses, Intraocular
Alternative methods
Phacoemulsification
business.industry
Significant difference
Retinal Detachment
Retinal detachment
Axial length
medicine.disease
United Kingdom
Sensory Systems
Confidence interval
030104 developmental biology
030221 ophthalmology & optometry
Optometry
Surgery
business
Subjects
Details
- ISSN :
- 08863350
- Volume :
- 42
- Database :
- OpenAIRE
- Journal :
- Journal of Cataract and Refractive Surgery
- Accession number :
- edsair.doi.dedup.....7b3a8763781462e06d8b49eca7481740