Back to Search Start Over

Control of energy absorption rate in transmission line radiofrequency exposure systems

Authors :
Dorothy J. Elliott
Joseph S. Ali
Source :
Bioelectromagnetics. 8:165-172
Publication Year :
1987
Publisher :
Wiley, 1987.

Abstract

A frequent problem in the radiofrequency (RF) irradiation of experimental animals in health effects studies is the temporal variation of the specific absorption rate (SAR) with animal movement. An RF power controller that regulates the energy absorption rate has been designed for use with transmission line exposure systems that utilize the power difference method to monitor the SAR. The controller operates by altering the incident power to the exposure cell in order to compensate for the change in RF energy absorption rate that is due to animal motion. A circuit diagram is presented as well as experimental data under three conditions of exposure. The controller is effective in maintaining the mean value of energy absorption rate at the setpoint value even for the case of a highly active animal.

Details

ISSN :
1521186X and 01978462
Volume :
8
Database :
OpenAIRE
Journal :
Bioelectromagnetics
Accession number :
edsair.doi.dedup.....7605f658db7d2adab59639e4bf1324a1
Full Text :
https://doi.org/10.1002/bem.2250080207