Evaluating cross contamination on a shared ventilator
- Source :
- Emergency Medicine Journal : EMJ
- Publication Year :
- 2020
- Publisher :
- BMJ Publishing Group, 2020.
-
Abstract
- BackgroundDisasters have the potential to cause critical shortages of life-saving equipment. It has been postulated that during patient surge, multiple individuals could be maintained on a single ventilator. This was supported by a previous trial that showed one ventilator could support four sheep. The goal of our study is to investigate if cross contamination of pathological agents occurs between individuals on a shared ventilator with strategically placed antimicrobial filters.MethodsA multipatient ventilator circuit was assembled using four sterile, parallel standard tubing circuits attached to four 2 L anaesthesia bags, each representing a simulated patient. Each ‘patient’ was attached to a Heat and Moisture Exchange filter. An additional bacterial/viral filter was attached to each expiratory limb. ‘Patient-Lung’ number 1 was inoculated with an isolate of Serratia marcescens, and the circuit was run for 24 hours. Each ‘lung’ and three points in the expiratory limb tubing were washed with broth and cultured. All cultures were incubated for 48 hours with subcultures performed at 24 hours.ResultsWashed cultures of patient 2, 3 and 4 failed to demonstrate growth of S. marcescens. Cultures of the distal expiratory tubing, expiratory limb connector and expiratory limb prefilter tubing yielded no growth of S. marcescens at 24 or 48 hours.ConclusionBased on this circuit configuration, it is plausible to maintain four individuals on a single ventilator for 24 hours without fear of cross contamination.
- Subjects :
- safety
Ventilator circuit
Expiratory limb
030231 tropical medicine
resuscitation
Economic shortage
Critical Care and Intensive Care Medicine
infectious diseases
03 medical and health sciences
0302 clinical medicine
Intensive care
Medicine
Humans
030212 general & internal medicine
Original Research
intensive care
Cross Infection
Ventilators, Mechanical
business.industry
ventilation
General Medicine
Equipment Design
Contamination
Viral filter
Anesthesia
Emergency Medicine
Equipment Contamination
business
Filtration
Subjects
Details
- Language :
- English
- ISSN :
- 14720213 and 14720205
- Database :
- OpenAIRE
- Journal :
- Emergency Medicine Journal : EMJ
- Accession number :