Back to Search Start Over

Technical note: The effect of the 4-mm-collimator output factor on gamma knife dose distributions

Authors :
Burke Dial
Robert Clark
Hassaan Alkhatib
William Neglia
Dimitris Mihailidis
Raleigh Boulware
Curtis Worthington
John P. Gibbons
Source :
Journal of Applied Clinical Medical Physics
Publication Year :
2003
Publisher :
Wiley, 2003.

Abstract

We present results of investigations of the clinical significance of variations in the value of the 4-mm-collimator output factor, OF4/18. Changes in treatment volume, dose-volume histograms (DVHs), and isodose distributions were studied, by varying OF4/18 up to 20. The variations were performed on a sample of clinical patient treatment plans for which the 4 mm collimator was used. Although smaller effects are noted for the prescription isodose line, greater dosimetric changes occur for higher dose regions within the target. 2003 American College of Medical Physics.

Details

ISSN :
15269914
Volume :
4
Database :
OpenAIRE
Journal :
Journal of Applied Clinical Medical Physics
Accession number :
edsair.doi.dedup.....75d1a79189bc4e186e4e4b536ae4c352
Full Text :
https://doi.org/10.1120/jacmp.v4i4.2512