Back to Search Start Over

Precise and fast control of the dissolved oxygen level for tumor-on-chip

Authors :
Charlotte Bouquerel
William César
Lara Barthod
Sarah Arrak
Aude Battistella
Giacomo Gropplero
Fatima Mechta-Grigoriou
Gérard Zalcman
Maria Carla Parrini
Marine Verhulsel
Stéphanie Descroix
Source :
Lab on a chip. 22(22)
Publication Year :
2022

Abstract

Oxalis features: independent control of pO2, pH and the liquid flowrate. pO2 equilibration time in the medium: 3 minutes. pO2 accuracy: 3 mmHg. Flowrate as low as 1 μL min−1 to avoid shear stress.