Investigation of the Accuracy of Four Intraoral Scanners in Mandibular Full-Arch Digital Implant Impression: A Comparative In Vitro Study
- Source :
- International Journal of Environmental Research and Public Health; Volume 19; Issue 8; Pages: 4719
- Publication Year :
- 2022
- Publisher :
- MDPI AG, 2022.
-
Abstract
- Background: We compare the accuracy of new intraoral scanners (IOSs) in full-arch digital implant impressions. Methods: A master model with six scan bodies was milled in poly(methyl methacrylate), measured by using a coordinate measuring machine, and scanned 15 times with four IOSs: PrimeScan, Medit i500, Vatech EZ scan, and iTero. The software was developed to identify the position points on each scan body. The 3D position and distance analysis were performed. Results: The average and ± standard deviation of the 3D position analysis was 29 μm ± 6 μm for PrimeScan, 39 μm ± 6 μm for iTero, 48 μm ± 18 μm for Mediti500, and 118 μm ± 24 μm for Vatech EZ scan (p < 0.05). Conclusions: All IOSs are able to make a digital complete implant impression in vitro according to the average misfit value reported in literature (150 μm); however, the 3D distance analysis showed that only the Primescan and iTero presented negligible systematic error sources.
Details
- ISSN :
- 16604601
- Volume :
- 19
- Database :
- OpenAIRE
- Journal :
- International Journal of Environmental Research and Public Health
- Accession number :
- edsair.doi.dedup.....72ffe054c6a5423a5b74eba7739f4251
- Full Text :
- https://doi.org/10.3390/ijerph19084719