Back to Search Start Over

Dilemma of Membrane Biocompatibility and Reuse

Authors :
J. Vienken
Aileen Grassmann
Horst Klinkmann
Source :
Artificial Organs. 20:426-432
Publication Year :
1996
Publisher :
Wiley, 1996.

Abstract

Numerous articles have been published on the multiple use of dialyzers and on the effect of different reprocessing chemicals and techniques on the dialyzer biocompatibility and performance. The results often appear contradictory, especially those comparing standard biocompatibility parameters. Despite this confusion, a discerning review of the published works allows certain limited conclusions to be drawn. Reprocessing of used hemodialyzers changes the biocompatibility profile of a dialyzer as defined by the parameters complement activation. leukopenia, and cytokine release. The effect of reprocessing depends on the chemicals and reprocessing technique applied and also on the type of membrane polymer being subjected to the reprocessing procedure. Reports of pyrogenic reactions indicate that the flux of the membrane also influences how suitable it is for safe reuse. An increased risk of allergic and pyrogenic reactions appears to be associated with dialyzer reuse. Furthermore, there has been a lack of investigations into the immunologic effect of the layer of adsorbed and chemically altered proteins that remains on the inner surface of reprocessed dialyzers. We conclude that the clinical benefit of dialyzer reuse cannot be generally accepted from a biocompatibility point of view.

Details

ISSN :
15251594 and 0160564X
Volume :
20
Database :
OpenAIRE
Journal :
Artificial Organs
Accession number :
edsair.doi.dedup.....70ca5f4c2f840c370ee08ac3b1876732
Full Text :
https://doi.org/10.1111/j.1525-1594.1996.tb04527.x