Back to Search
Start Over
Effects of patient orientations, landmark positions, and device positions on the MRI RFâinduced heating for modular external fixation devices
- Source :
- Magnetic Resonance in Medicine. 85:1669-1680
- Publication Year :
- 2020
- Publisher :
- Wiley, 2020.
-
Abstract
- Purpose This paper studies the RF-induced heating for modular external fixation devices applied on the leg regions of the human bodies. Through numerical investigations of RF-induced heating related to different patient orientations, landmark positions, and device positions under 1.5T and 3T MRI systems, simple and practical methods to reduce RF-induced heating are recommended. Methods Numerical simulations using a full-wave electromagnetic solver based on the finite-difference time-domain method were performed to characterize the effects of patient orientations (head-first/feet-first), landmark positions (the scanning area of the patient), and device positions (device on left or right leg) on the RF-induced heating of the external fixation devices. The G32 coil design and three anatomical human models (Duke model, Ella model, and Fats model) were adopted to model the MRI RF coil and the patients. Results The relative positions of the patient, device, and coil can significantly affect the RF-induced heating. With other conditions remaining the same, changing the device position or patient orientation can lead to a peak 1-g averaged spatial absorption ratio variation of a factor around four. By changing the landmark position and the patient orientation, the RF-induced heating can be reduced from 1323.6 W/kg to 217.5 W/kg for the specific scanning situations studied. Conclusion Patient orientations, landmark positions, and device positions influence the RF-induced heating of modular external fixation devices at 1.5 T and 3 T. These features can be used to reduce the RF-induced heating during MRI simply and practically.
- Subjects :
- Hot Temperature
Materials science
External Fixators
Radio Waves
Acoustics
030218 nuclear medicine & medical imaging
Heating
03 medical and health sciences
0302 clinical medicine
Fracture Fixation
Position (vector)
Humans
Radiology, Nuclear Medicine and imaging
Lead (electronics)
Landmark
Phantoms, Imaging
business.industry
Patient orientation
External fixation devices
Modular design
Magnetic Resonance Imaging
Electromagnetic coil
business
psychological phenomena and processes
030217 neurology & neurosurgery
Radiofrequency coil
Subjects
Details
- ISSN :
- 15222594 and 07403194
- Volume :
- 85
- Database :
- OpenAIRE
- Journal :
- Magnetic Resonance in Medicine
- Accession number :
- edsair.doi.dedup.....6ec5b6474832a1616cdd12bd06f4bae2
- Full Text :
- https://doi.org/10.1002/mrm.28514