Trueness of ten intraoral scanners in determining the positions of simulated implant scan bodies

Authors :
Ryan Jin-Young Kim
Ji-Man Park
Goran I. Benic
University of Zurich
Park, Ji-Man
Source :
Scientific Reports, Scientific Reports, Vol 11, Iss 1, Pp 1-9 (2021)
Publication Year :
2021
Publisher :
Springer Science and Business Media LLC, 2021.

Abstract

Few investigations have evaluated the 3-dimensional (3D) accuracy of digital implant scans. The aim of this study was to evaluate the performance of 10 intraoral scanners (IOSs) (CEREC Omnicam, CEREC Primescan, CS 3600, DWIO, i500, iTero Element, PlanScan, Trios 2, Trios 3, and True Definition) in obtaining the accurate positions of 6 cylinders simulating implant scan bodies. Digital scans of each IOS were compared with the reference dataset obtained by means of a coordinate measuring machine. Deviation from the actual positions of the 6 cylinders along the XYZ axes and the overall 3D deviation of the digital scan were calculated. The type of IOSs and position of simulated cylindrical scan bodies affected the magnitude and direction of deviations on trueness. The lowest amount of deviation was found at the cylinder next to the reference origin, while the highest deviation was evident at the contralateral side for all IOSs (p p > 0.05), and the DWIO and PlasScan had the lowest trueness in partially edentulous mandible digital implant scans (p