Specificity of B-Type Natriuretic Peptide Assays: Cross-Reactivity with Different BNP, NT-proBNP, and proBNP Peptides

Authors :
Fred S. Apple
Jens P. Goetze
Ranka Ler
Allan S. Jaffe
Jordi Ordóñez-Llanos
Amy K. Saenger
Olaia Rodríguez-Fraga
Source :
Clinical Chemistry. 63:351-358
Publication Year :
2017
Publisher :
Oxford University Press (OUP), 2017.

Abstract

BACKGROUNDB-type natriuretic peptides (BNPs) are used clinically to diagnose and monitor heart failure and are present in the circulation as multiple proBNP-derived fragments. We investigated the specificity of BNP immunoassays with glycosylated and nonglycosylated BNP, N-terminal proBNP (NT-proBNP), and proBNP peptides to probe the cross-reactivity of each assay.METHODSNine B-type natriuretic peptides were studied,including synthetic and recombinant BNP (Shionogi, Scios, Mayo), human and synthetic glycosylated and nonglycosylated NT-proBNP (HyTest, Roche Diagnostics), and human glycosylated and nonglycosylated proBNP (HyTest, Scios). Five BNP [Abbott, Abbott POC, Alere, Beckman Coulter, Siemens (Centaur)], 9 NT-proBNP [Ortho-Clinical Diagnostics, Roche, Response, bioMerieux, Siemens (Dimension, Immulite, Stratus CS), Mitsubishi] and 3 research-use-only proBNP immunoassays [Biosite (Alere), Bio-Rad, Goetze] were evaluated. Specificity was assessed by calculating the recovery between baseline and peptide-spiked human plasma pools at target concentrations of 100 ng/L BNP, 300 ng/L proBNP, or 450 ng/L NT-proBNP. All assays were performed in duplicate.RESULTSBNP and NT-proBNP assays demonstrated substantial cross-reactivity with proBNP peptides. NT-proBNP assays do not detect glycosylated forms of either NT-proBNP or proBNP. proBNP assays preferentially detect the BNP 1–32 peptide and have minimal cross-reactivity with BNP peptides and glycosylated proBNP.CONCLUSIONSBNP or NT-proBNP results are not transferable among the current existing immunoassays owing to their differences in cross-reactivity and ability to detect various glycosylated forms of proBNP-derived fragments. Opportunities remain to standardize and harmonize BNP and NT-proBNP assays, as well as to develop specific proBNP assays, to widen their clinical scope of use.

Details

ISSN :
15308561 and 00099147
Volume :
63
Database :
OpenAIRE
Journal :
Clinical Chemistry
Accession number :
edsair.doi.dedup.....6a530f48e669bc873a55cadc1f44aac8