Back to Search
Start Over
Quality assurance of Mo-99/Tc-99m radionuclide generators
- Source :
- Acta Scientifica Naturalis, Vol 5, Iss 1, Pp 40-47 (2018)
- Publication Year :
- 2018
- Publisher :
- Walter de Gruyter GmbH, 2018.
-
Abstract
- Gamma-ray spectrometry analyses of the radionuclide content of eluate from two Mo-99/Tc-99m radionuclide generators POLTECHNET have been performed. The relative activities of 99Mo 103Ru and 131I radioisotopes with respect to the activity of 99mTc at different time intervals after the primary pertechnetate elution of the generators have been analyzed. The relative activities of the isotopes were determined and compared to the radionuclidic purity requirements for 99mTc.
- Subjects :
- Pulmonary and Respiratory Medicine
Physics
Geography (General)
Technetium-99m generator
business.industry
Science
Radiochemistry
quality assurance
02 engineering and technology
030204 cardiovascular system & hematology
021001 nanoscience & nanotechnology
03 medical and health sciences
0302 clinical medicine
gamma-ray spectroscopy
G1-922
nuclear medicine
Pediatrics, Perinatology, and Child Health
Radionuclide Generator
0210 nano-technology
business
technetium generator
Quality assurance
Subjects
Details
- ISSN :
- 23675144
- Volume :
- 5
- Database :
- OpenAIRE
- Journal :
- Acta Scientifica Naturalis
- Accession number :
- edsair.doi.dedup.....69df076a84cd0da5484477a522632262
- Full Text :
- https://doi.org/10.2478/asn-2018-0006