Double Guidewire Method: A Novel Technique for Correction of Migrated Tenckhoff Peritoneal Dialysis Catheter
- Source :
- Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis. 23:587-590
- Publication Year :
- 2003
- Publisher :
- SAGE Publications, 2003.
-
Abstract
- Twenty-two consecutive patients with a continuous ambulatory peritoneal dialysis (CAPD) catheter malfunctioning due to catheter migration were treated with a novel radiological manipulation technique, the “double guidewire method.” The first guidewire is used to correct the direction of the catheter tip and the second wire is used to anchor the CAPD catheter so that an ideal course of the catheter can be maintained during removal of the first guidewire. Immediate catheter repositioning was achieved in 19 of 22 patients, and durable repositioning success was achieved in 13 patients. In conclusion, the “double guidewire method” is a simple but effective technique for prolonging CAPD catheter life in patients with malfunction due to catheter migration.
- Subjects :
- Adult
Male
Tenckhoff catheter
Novel technique
medicine.medical_specialty
medicine.medical_treatment
030232 urology & nephrology
Catheterization
Peritoneal dialysis
03 medical and health sciences
0302 clinical medicine
Foreign-Body Migration
medicine
Peritoneal dialysis catheter
Humans
030212 general & internal medicine
Aged
business.industry
Continuous ambulatory peritoneal dialysis
General Medicine
Middle Aged
medicine.disease
Surgery
Catheter
Nephrology
Female
Radiology
Dialisis peritoneal
business
Peritoneal Dialysis
Kidney disease
Subjects
Details
- ISSN :
- 17184304 and 08968608
- Volume :
- 23
- Database :
- OpenAIRE
- Journal :
- Peritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis
- Accession number :
- edsair.doi.dedup.....6932b56bd169ca94566c018585dfab19
- Full Text :
- https://doi.org/10.1177/089686080302300612