Back to Search Start Over

Quantification of polychlorinated dibenzo-p-dioxins and dibenzofurans by direct injection of sample extract into the comprehensive multidimensional gas chromatograph/high-resolution time-of-flight mass spectrometer

Authors :
Shibata Yasuyuki
Ubukata Masa-aki
Tanabe Kiyoshi
Anezaki Katsunori
Otsuka Hideyuki
Ito Hiroyasu
Tanaka Kazuo
Kusai Akihiko
Takazawa Yoshikatsu
Hashimoto Shunji
Fushimi Akihiro
Source :
Journal of chromatography. A. 1178(1-2)
Publication Year :
2007

Abstract

Polychlorinated dibenzo-p-dioxins and dibenzofurans in crude extracts of fly ash and flue gas from municipal waste incinerators were quantified using a comprehensive multidimensional gas chromatograph (GC x GC) coupled to a high-resolution time-of-flight mass spectrometer (HR-TOFMS). For identification and quantification, we developed our own program to prepare 3D chromatograms of selected mass numbers from the data of the GC x GC/HR-TOFMS. Isolation of all congeners with a TCDD toxic equivalency factor from the other isomers by only one injection was confirmed. The instrumental detection limit of TCDD on the GC x GC/HR-TOFMS was 0.9 pg by the relative calibration method. Quantification of these substances in the crude extracts was achieved by direct injection to the GC x GC/HR-TOFMS. The results agree with the values obtained using a generic gas chromatography/high-resolution mass spectrometry (GC/HRMS) system. It was confirmed that measurement by high-resolution TOFMS and GC x GC effectively reduces interference from other chemicals.

Details

ISSN :
00219673
Volume :
1178
Issue :
1-2
Database :
OpenAIRE
Journal :
Journal of chromatography. A
Accession number :
edsair.doi.dedup.....686294cb49f2fd72fb84c6a1e50d3ac8