Back to Search Start Over

A bidirectional load applicator for the investigation of skin response to mechanical stress

Authors :
Joseph L. Garbini
M.S. Allen
Joan E. Sanders
J.E. Jorgensen
J.M. Leschen
Source :
IEEE Transactions on Biomedical Engineering. 44:290-296
Publication Year :
1997
Publisher :
Institute of Electrical and Electronics Engineers (IEEE), 1997.

Abstract

Instrumentation was developed to apply controlled biaxial (normal and shear) forces to the skin of a human or animal subject. The instrument mimicked any reference waveform within the constraints of a bandwidth of 15 Hz, a maximum force of 20 N, and displacement ranges of 15 mm for the normal direction and 18 mm for the shear direction. Two shaker motors, positioned with their axes parallel, were used with a low effective mass linkage and small-angle rotational joints to deliver the force. A digital feedback controller independently controlled the instantaneous normal and shear forces and recorded the resultant displacements. Evaluations on human and animal (pig) subjects demonstrated mean absolute errors between the applied and reference waveforms of less than 1.2% full-scale output for both the normal and shear directions. No degradation in performance was apparent over the course of a 1-h loading session. The instrument is to be used for the investigation of skin adaptation to mechanical stress, information that could be used to design new therapeutic methods to encourage skin load-tolerance.

Details

ISSN :
00189294
Volume :
44
Database :
OpenAIRE
Journal :
IEEE Transactions on Biomedical Engineering
Accession number :
edsair.doi.dedup.....6761a1893da89d8d260df97c71ff36a6