Multicenter Experience Evaluating Transcatheter Pulmonary Valve Replacement in Bovine Jugular Vein (Contegra) Right Ventricle to Pulmonary Artery Conduits

Authors :
Saar Danon
Brian H. Morray
Mark Hoyer
Younes Boudjemline
Dennis W. Kim
Aimee K. Armstrong
Darren P. Berman
Doff B. McElhinney
Marc Gewillig
Mary Hunt Martin
Athar M. Qureshi
Jeffrey W. Delaney
Jeffery Meadows
Henri Justino
Thomas K. Jones
Martin L. Bocks
Elena K. Grant
Source :
Circulation. Cardiovascular interventions. 10(6)
Publication Year :
2016

Abstract

Background— Follow-up of transcatheter pulmonary valve replacement (TPVR) with the Melody valve has demonstrated good short-term and long-term outcomes, but there are no published studies focused on valve performance in the Contegra bovine jugular vein conduit. Methods and Results— This is a retrospective, multicenter study of the short- and intermediate-term outcomes of Melody TPVR within the Contegra conduit in the right ventricle to pulmonary artery position. Data from 13 centers were included in the analysis. During the study period, 136 patients underwent 139 catheterizations for attempted Melody TPVR with a median follow-up of 3 years (1 day to 9.1 years). Of the 136 patients, 117 underwent successful Melody TPVR. Two patients underwent a second Melody TPVR. The majority of patients underwent placement of ≥1 stents before transcatheter pulmonary valve implantation. There was a significant reduction in peak conduit pressure gradient acutely after transcatheter pulmonary valve implantation (39 versus 10 mm Hg; P P Conclusions— Melody TPVR in Contegra conduits is safe and effective and can be performed in a wide range of conduit sizes with preserved valve function and low incidence of stent fracture and endocarditis.

Details

ISSN :
19417632
Volume :
10
Issue :
6
Database :
OpenAIRE
Journal :
Circulation. Cardiovascular interventions
Accession number :
edsair.doi.dedup.....64ba55cf8f9a4b56fedaeafcd91199fa