Back to Search Start Over

A comprehensive experimental characterization of the iPIX gamma imager

Authors :
V. Paradiso
P. Couturier
K. Amgarou
Frank Becker
J. Handley
N. Menaa
Audrey Patoz
F. Bonnet
Source :
Journal of Instrumentation. 11:P08012-P08012
Publication Year :
2016
Publisher :
IOP Publishing, 2016.

Abstract

The results of more than 280 different experiments aimed at exploring the main features and performances of a newly developed gamma imager, called iPIX, are summarized in this paper. iPIX is designed to quickly localize radioactive sources while estimating the ambient dose equivalent rate at the measurement point. It integrates a 1 mm thick CdTe detector directly bump-bonded to a Timepix chip, a tungsten coded-aperture mask, and a mini RGB camera. It also represents a major technological breakthrough in terms of lightness, compactness, usability, response sensitivity, and angular resolution. As an example of its key strengths, an (241)Am source with a dose rate of only few nSv/h can be localized in less than one minute.

Details

ISSN :
17480221
Volume :
11
Database :
OpenAIRE
Journal :
Journal of Instrumentation
Accession number :
edsair.doi.dedup.....61ddf8b7b261afd876f973359d7be984
Full Text :
https://doi.org/10.1088/1748-0221/11/08/p08012