Back to Search Start Over

Distortion of the CoreValve during transcatheter aortic valve-in-valve implantation due to valve dislocation

Authors :
Jean-Claude Laborde
Géraud Souteyrand
Pascal Motreff
Krzysztof Wilczek
Nicolas Durel
Jean-René Lusson
Pascal Chabrot
Andrea Innorta
Piotr Chodór
Lionel Camilleri
CHU Clermont-Ferrand
Service de chirurgie cardiaque
Institut Pascal (IP)
SIGMA Clermont (SIGMA Clermont)-Centre National de la Recherche Scientifique (CNRS)-Université Clermont Auvergne [2017-2020] (UCA [2017-2020])
Service de Cardiologie Maladies Vasculaires [CHU Clermont-Ferrand]
CHU Gabriel Montpied [Clermont-Ferrand]
CHU Clermont-Ferrand-CHU Clermont-Ferrand
IRSN-DSU-SERAC
Laboratory for the Experimental Study of Containment, Air Cleaning and Ventilation
Image Science for Interventional Techniques (ISIT)
Université d'Auvergne - Clermont-Ferrand I (UdA)-Clermont Université-Centre National de la Recherche Scientifique (CNRS)
SIGMA Clermont (SIGMA Clermont)-Université Clermont Auvergne [2017-2020] (UCA [2017-2020])-Centre National de la Recherche Scientifique (CNRS)
Service d'études et de recherches en aérocontamination et en confinement (IRSN/DSU/SERAC)
Institut de Radioprotection et de Sûreté Nucléaire (IRSN)
Source :
Cardiovascular Revascularization Medicine, Cardiovascular Revascularization Medicine, Elsevier, 2013, 14 (5), pp.294-298. ⟨10.1016/j.carrev.2013.05.004⟩, Cardiovascular Revascularization Medicine, 2013, 14 (5), pp.294-298. ⟨10.1016/j.carrev.2013.05.004⟩
Publication Year :
2013
Publisher :
Elsevier BV, 2013.

Abstract

Nowadays transcatheter aortic valve implantation (TAVI) is an accepted alternative to surgical aortic valve replacement for high-risk patients (pts). Successful TAVI procedures for failed aortic surgical bioprosthesis (TAV-in-SAV) have already been reported. In the presented two cases of TAV-in-SAV implantation a strut distortion of the stent was revealed on angiographic imaging and confirmed on control CT scan. In both procedures, a dislocation of the medtronic core valve (MCV) prosthesis during implantation led to valve retrieval, with a necessity of reloading it in the 18F introducer before subsequent implantation of the same valve in correct position.