Decreased Detection of Hypovascular Liver Tumors With MDCT in Obese Patients: A Phantom Study

Authors :
Rendon C. Nelson
Sebastian T. Schindera
Zsolt Szucs-Farkas
Daniele Marin
Hanno Hoppe
Jaled Charimo Torrente
Thomas D. Ruder
Source :
American Journal of Roentgenology. 196:W772-W776
Publication Year :
2011
Publisher :
American Roentgen Ray Society, 2011.

Abstract

The purpose of this article is to assess the impact of large patient size on the detection of hypovascular liver tumors with MDCT and the effect of a noise filter on image quality and lesion detection in obese patients.A liver phantom with 45 hypovascular tumors (diameters of 5, 10, and 15 mm) was placed into two water containers mimicking intermediate and large patients. The containers were scanned with a 64-MDCT scanner. The CT dataset from the large phantom was postprocessed using a noise filter. The image noise was measured and the contrast-to-noise ratio (CNR) of the tumors was calculated. Tumor detection was independently performed by three radiologists in a blinded fashion.The application of the noise filter in the large phantom yielded a reduction of image noise by 42% (p0.0001). The CNR values of the tumors in the nonfiltered and filtered large phantom were lower than that in the intermediate phantom (p0.05). In the non-filtered and filtered large phantom, 25% and 19% fewer tumors, respectively, were detected on average compared with the intermediate phantom (p0.01).The risk of missing hypovascular liver tumors with CT is substantially increased in large patients. A noise filter improves image quality in obese patients.

Details

ISSN :
15463141 and 0361803X
Volume :
196
Database :
OpenAIRE
Journal :
American Journal of Roentgenology
Accession number :
edsair.doi.dedup.....60f443c2f1b6be62c2ed186db1ca2f29