Back to Search Start Over

How to maximize QRS narrowing

Authors :
Olivia Braun
Julia W. Erath
Stefan H. Hohnloser
Mate Vamos
Source :
Herzschrittmachertherapie + Elektrophysiologie. 30:229-232
Publication Year :
2019
Publisher :
Springer Science and Business Media LLC, 2019.

Abstract

Cardiac resynchronization therapy (CRT) resulting in maximal QRS narrowing may be associated with improved outcomes. Various atrioventricular (AV) delay settings, including the new SyncAV™ algorithm (St. Jude Medical/Abbott, St. Paul, MN, USA), aimed at maximal QRS narrowing were tested in an 81-year old CRT recipient. Maximal QRS narrowing from 160 to 100 ms was achieved with a manually programmed value of SyncAV™ −30 ms. At 2 months, the patient proved to be a CRT super-responder. SyncAV™ algorithm is a new way for effective QRS narrowing with potentially improved outcomes.

Details

ISSN :
14351544 and 09387412
Volume :
30
Database :
OpenAIRE
Journal :
Herzschrittmachertherapie + Elektrophysiologie
Accession number :
edsair.doi.dedup.....6070e3563dc1f408c5956880b8680fb6