Back to Search Start Over

A sampler prototype for the simultaneous collection of exhaled air and breath condensate

Authors :
Denise Biagini
Silvia Ghimenti
Tommaso Lomonaco
S. Antoni
Francesca G. Bellagambi
Pietro Salvo
F. Di Francesco
Roger Fuoco
Source :
EMBC
Publication Year :
2019
Publisher :
IEEE, 2019.

Abstract

Exhaled air and breath condensate contain a large number of health biomarkers, such as volatile and semi-volatile organic compounds, proteins and lipids. Nowadays, the collection of breath samples is carried out by commercial or lab-made sampling systems that collect only one type of sample (e.g. gaseous or condensate phase), thus limiting the diagnostic capability of breath tests. This work presents a portable prototype optimized for the simultaneous collection of gaseous exhaled breath and exhaled breath condensate within five minutes. The system is fully portable and has a total weight of about 1 Kg. An illustrative determination of ethanol, isoprene, acetone, isopropyl alcohol, 1-propanol, 2-butanone, 2-pentanone, toluene and xylenes in breath, and cortisol and 8-iso-prostaglandin F 2α in breath condensate is discussed.

Details

Database :
OpenAIRE
Journal :
2019 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)
Accession number :
edsair.doi.dedup.....5e1efe759a98100293819aa6160ee921