Screening for Previous Refractive Surgery in Eye Bank Corneas by Using Optical Coherence Tomography

Authors :
Andrew M. Rollins
Yan Li
David Huang
Maolong Tang
Roger C. Lin
Marcy McLain
Joseph A. Izatt
Source :
Cornea. 26:594-599
Publication Year :
2007
Publisher :
Ovid Technologies (Wolters Kluwer Health), 2007.

Abstract

PURPOSE To use optical coherence tomography (OCT) to detect previous refractive surgery in donor corneas. METHODS We constructed a tabletop OCT scanner operating at 1310-nm wavelength. Donor corneas at the Cleveland Eye Bank were scanned while sealed within the sterile container immersed in Optisol GS. OCT scanning was performed with 7.6-mm-long lines (512 axial scans) along 8 meridians. Anterior and posterior corneal surfaces were automatically mapped using image processing software that we developed. Curvature was computed from the best parabolic fit in the central 5-mm diameter. Layered analysis of the stromal reflectivity was also performed. Twenty-nine corneas from 19 donors were examined. Five had a history of laser in situ keratomileusis (LASIK). RESULTS The flap interfaces could not be visualized on slit-lamp or OCT images but were confirmed by histology. The death-to-scan time was 22.1 +/- 11.4 (SD) hours for normal corneas and 100.6 +/- 57.5 hours for LASIK corneas. The anterior surface power was 67.5 +/- 2.5 D in control corneas and 64.5 +/- 2.4 D in LASIK corneas (P = 0.023). There was no significance between the 2 groups in terms of posterior curvature and thickness parameters. The anterior/posterior reflectivity ratio in the central 4-mm diameter was significantly lower in post-LASIK corneas than in control (P < 0.05). CONCLUSIONS OCT provides thickness, topography, and reflectivity maps of donor corneas without taking them out of preservation medium and container. The anterior curvature and the anterior/posterior stromal reflectivity ratio may be useful for detecting previous LASIK.

Details

ISSN :
02773740
Volume :
26
Database :
OpenAIRE
Journal :
Cornea
Accession number :
edsair.doi.dedup.....5ae4fb420cbbcd02f84544ae58461171
Full Text :
https://doi.org/10.1097/ico.0b013e31803c5535