Back to Search Start Over

Measurements of Ankle Dorsiflexion in Stroke Subjects obtained using standardised dorsiflexion force

Authors :
Jennifer Lyn Keating
Merilyn Mackenzie
Claire Parks
Publication Year :
2023
Publisher :
La Trobe, 2023.

Abstract

This study investigated the reliability of measurements of ankle dorsiflexion obtained using the Lidcombe Template, an instrument that allows the magnitude and direction of force applied to dorsiflex the foot to be measured and standardised. Ten unimpaired physiotherapy students and 21 subjects who had suffered stroke were tested twice. Twenty minutes separated tests. Measurements of passive dorsiflexion range were highly reliable for both groups (r > 0.92) when the mean of three measurements was used. Significantly more variability occurred in measurements of impaired subjects than unimpaired subjects. For the subjects tested, for 95 per cent confidence that real differences exist between measurements taken 20 minutes apart, 7 degrees and 3 degrees must be allowed around measurements of impaired and unimpaired subjects respectively. These small error margins confer confidence in the potential utility of this instrument for measuring ankle dorsiflexion.

Details

Database :
OpenAIRE
Accession number :
edsair.doi.dedup.....5a50b25c076f8e5df1e37ccacaa77b82
Full Text :
https://doi.org/10.26181/22199797