Evaluation of the Technicon Immuno I random access immunoassay analyzer and calculation of pediatric reference ranges for endocrine tests, T-uptake, and ferritin
- Source :
- Clinical biochemistry. 28(2)
- Publication Year :
- 1995
-
Abstract
- Objectives : Evaluation of the precision, accuracy, and user-friendliness of the Technician Immuno I. Calculation of pediatric reference ranges for ferritin and endocrine tests run on Immuno I. Methods : Precision and accuracy were measured using controls and method comparison studies. Pediatric reference ranges were calculated by comparing the Immuno I results for 100 patients with those of the Abbott IMx and TDx. The regression equation obtained was then used to convert the [Mx and TDx reference ranges to reference ranges for the Immuno I. Results : The Immuno I provided both accurate and precise measurement of drugs and endocrine hormones. Pediatric reference ranges were obtained for ferritin and all endocrine tests. Conclusion : The Immuno I is user-friendly and provides reliable measurement of both the drugs tested and endocrine tests on a micro-sample. Reagent and curve stability are excellent.
- Subjects :
- Adult
Male
Accuracy and precision
Spectrum analyzer
Pathology
medicine.medical_specialty
Adolescent
Clinical Biochemistry
AutoAnalyzer
User-Computer Interface
medicine
Humans
Child
Reference standards
Immunoassay
Endocrine Test
Chromatography
Autoanalysis
medicine.diagnostic_test
biology
Chemistry
Infant, Newborn
Infant
Reproducibility of Results
General Medicine
Reference Standards
Hormones
Ferritin
Thyroxine
Method comparison
Evaluation Studies as Topic
Child, Preschool
Ferritins
biology.protein
Female
Subjects
Details
- ISSN :
- 00099120
- Volume :
- 28
- Issue :
- 2
- Database :
- OpenAIRE
- Journal :
- Clinical biochemistry
- Accession number :
- edsair.doi.dedup.....58db9f03d90f1429bfc61768668e86eb