Back to Search
Start Over
ESTIMATION OF ENTRANCE SURFACE AIR KERMA IN DIGITAL RADIOGRAPHIC EXAMINATIONS
- Source :
- Radiation Protection Dosimetry. 193:16-23
- Publication Year :
- 2021
- Publisher :
- Oxford University Press (OUP), 2021.
-
Abstract
- Purpose: Contribution of radiation doses from medical X-ray examination to collective dose is significant. Unusually, high doses may increase the risk of stochastic effects of radiations. Therefore, radiation dose assessment was performed in 241 digital X-ray examinations in the study and was compared with published dose reference levels (DRLs). Methods: Entrance surface air kerma (ESAK) was calculated in chest PA, cervical AP/Lat, abdomen AP, lumbar AP/Lat and pelvis AP digital radiographic examinations (119 male and 122 female) following the International Atomic Energy Agency recommended protocol. Initially, 270 digital examinations were selected, reject analysis was performed and final 241 examinations were enrolled in the study for dose calculations. The exposure parameters and X-ray tube output were used for dose calculations. Effective doses were estimated with the help of conversion coefficients from ICRP 103. Results: Median ESAK (mGy) and associated effective doses obtained were cervical spine AP (1.30 mGy, 0.045 mSv), cervical spine Lat (0.25 mGy, 0.005 mSv), chest PA (0.11 mGy, 0.014 mSv), abdomen AP (0.90 mGy, 0.118 mSv), lumbar spine AP (1.52 mGy, 0.177 mSv), lumbar spine Lat (7.76 mGy, 0.209 mSv) and pelvis AP (0.82 mGy, 0.081 mSv). Results were compared with the studies of UK, Oman, India and Canada. Conclusion: The calculated ESAK and effective dose values were less than or close to previously published literature except for cervical spine AP and lumbar spine Lat. The results reinforce the need for radiation protection optimization, improving examination techniques and appropriate use of automatic exposure control in digital radiography. ESAK values reported in this study could further contribute to establishing local DRLs, regional DRLs and national DRLs.
- Subjects :
- Male
Canada
Oman
Radiography
India
Radiation Dosage
Effective dose (radiation)
030218 nuclear medicine & medical imaging
03 medical and health sciences
Kerma
Radiation Protection
0302 clinical medicine
Lumbar
Humans
Medicine
Radiology, Nuclear Medicine and imaging
Pelvis
Automatic exposure control
Digital radiography
Radiation
Radiological and Ultrasound Technology
business.industry
Public Health, Environmental and Occupational Health
General Medicine
Collective dose
medicine.anatomical_structure
030220 oncology & carcinogenesis
Female
business
Nuclear medicine
Subjects
Details
- ISSN :
- 17423406 and 01448420
- Volume :
- 193
- Database :
- OpenAIRE
- Journal :
- Radiation Protection Dosimetry
- Accession number :
- edsair.doi.dedup.....58b47112af260ddd9931facfa017d819
- Full Text :
- https://doi.org/10.1093/rpd/ncab018