DyNeuMo Mk-1: Design and pilot validation of an investigational motion-adaptive neurostimulator with integrated chronotherapy

Authors :
Mayela Zamora
Robert Toth
Francesca Morgante
Jon Ottaway
Tom Gillbe
Sean Martin
Guy Lamb
Tara Noone
Moaad Benjaber
Zachary Nairac
Devang Sehgal
Timothy G. Constandinou
Jeffrey Herron
Tipu Z. Aziz
Ivor Gillbe
Alexander L. Green
Erlick A.C. Pereira
Timothy Denison
Source :
Exp Neurol
Publication Year :
2022
Publisher :
Elsevier BV, 2022.

Abstract

There is growing interest in using adaptive neuro-modulation to provide a more personalized therapy experience that might improve patient outcomes. Current implant technology, however, can be limited in its adaptive algorithm capability. To enable exploration of adaptive algorithms with chronic implants, we designed and validated the, ‘Picostim DyNeuMo Mk-1’, (DyNeuMo Mk-1 for short), a fully-implantable, adaptive research stimulator that titrates stimulation based on circadian rhythms (e.g. sleep, wake) and the patient’s movement state (e.g. posture, activity, shock, free-fall). The design leverages off-the-shelf consumer technology that provides inertial sensing with low-power, high reliability, and relatively modest cost. The DyNeuMo Mk-1 system was designed, manufactured and verified using ISO 13485 design controls, including ISO 14971 risk management techniques to ensure patient safety, while enabling novel algorithms. The system was validated for an intended use case in movement disorders under an emergency-device authorization from the Medicines and Healthcare Products Regulatory Agency (MHRA). The algorithm configurability and expanded stimulation parameter space allows for a number of applications to be explored in both central and peripheral applications. Intended applications include adaptive stimulation for movement disorders, synchronizing stimulation with circadian patterns, and reacting to transient inertial events such as posture changes, general activity, and walking. With appropriate design controls in place, first-in-human research trials are now being prepared to explore the utility of automated motion-adaptive algorithms.

Details

ISSN :
00144886 and 10902430
Volume :
351
Database :
OpenAIRE
Journal :
Experimental Neurology
Accession number :
edsair.doi.dedup.....58722f86c4e0dc0ad5dda330979fe321
Full Text :
https://doi.org/10.1016/j.expneurol.2022.113977