A real-time sorting algorithm for in-beam PET of heavy-ion cancer therapy device

Authors :
Jiapeng Xu
Hong Su
Ke Lingyun
Changxu Pei
Xiuling Zhang
Jie Kong
Zuoqiao Yang
Jin-Da Chen
Yang Haibo
Pu Tianlei
Qian Yi
Zhao Hongyun
She Qianshun
Changxin Wang
Yan Junwei
Minchi Hu
Chengming Du
Source :
Nuclear Engineering and Technology, Vol 53, Iss 10, Pp 3406-3412 (2021)
Publication Year :
2021
Publisher :
Elsevier, 2021.

Abstract

A real-time digital time-stamp sorting algorithm used in the In-Beam positron emission tomography (In-Beam PET) is presented. The algorithm is operated in the field programmable gate array (FPGA) and a small amount of registers, MUX and memory cells are used. It is developed for sorting the data of annihilation event from front-end circuits, so as to identify the coincidence events efficiently in a large amount of data. In the In-Beam PET, each annihilation event is detected by the detector array and digitized by the analog to digital converter (ADC) in Data Acquisition Unit (DAQU), with a resolution of 14 bits and sampling rate of 50 MS/s. Test and preliminary operation have been implemented, it can perform a sorting operation under the event count rate up to 1 MHz per channel, and support four channels in total, count rate up to 4 MHz. The performance of this algorithm has been verified by pulse generator and 22Na radiation source, which can sort the events with chaotic order into chronological order completely. The application of this algorithm provides not only an efficient solution for selection of coincidence events, but also a design of electronic circuit with a small-scale structure.

Details

Language :
English
ISSN :
17385733
Volume :
53
Issue :
10
Database :
OpenAIRE
Journal :
Nuclear Engineering and Technology
Accession number :
edsair.doi.dedup.....57f77858d7fd5fecc4e409b7606e29a0