Impact of catheter down-sizing and power injector use on the amount of contrast medium delivered
- Source :
- Cardiovascular Intervention and Therapeutics. 25:24-28
- Publication Year :
- 2009
- Publisher :
- Springer Science and Business Media LLC, 2009.
-
Abstract
- The purpose of this study is to evaluate the effect of catheter down-sizing and power injector use on the amount of contrast medium delivered during diagnostic coronary angiography (CAG). We retrospectively studied 735 patients undergoing diagnostic CAG. Patients were divided into four groups according to catheter size [4 French (Fr) or 5 Fr] and the method of contrast injection (manual or power injection). Total contrast volume, procedure time, number of catheters used and angiogram quality were analyzed. Down-sizing catheters from 5 Fr to 4 Fr reduced the amount of contrast medium in both manual injection (from 153 ± 22 to 137 ± 17 ml, −10%, p
- Subjects :
- Coronary angiography
medicine.medical_specialty
medicine.diagnostic_test
business.industry
media_common.quotation_subject
Catheter size
Interventional radiology
Power injection
General Medicine
Injector
law.invention
Catheter
Contrast medium
law
Medicine
Contrast (vision)
Radiology, Nuclear Medicine and imaging
Radiology
Cardiology and Cardiovascular Medicine
business