An unusual mechanism causing inappropriate implantable cardioverter defibrillator shocks: transient reduction in R-wave amplitude
- Source :
- EP Europace. 9:694-696
- Publication Year :
- 2007
- Publisher :
- Oxford University Press (OUP), 2007.
-
Abstract
- Inappropriate shocks delivered by the implantable cardioverter defibrillator (ICD) are an increasingly recognized complication. The most frequent cause is related to supraventricular rhythms associated with fast conduction to the ventricles that are incorrectly detected as ventricular tachycardia leading to inappropriate antitachycardia pacing and/or shocks. Oversensing is a frequent cause of inappropriate shocks usually due to increased amplitude of the T-wave secondary to ischaemia or electrolyte disorders that lead to T-wave double counting. We describe an unusual case of T-wave double counting during sinus rhythm caused by transient reduction in R-wave amplitude with no changes in T-wave amplitude resulting in inappropriate shocks.
- Subjects :
- Male
Cardiac Complexes, Premature
medicine.medical_specialty
Unusual case
business.industry
medicine.medical_treatment
Implantable cardioverter-defibrillator
Ventricular tachycardia
medicine.disease
Defibrillators, Implantable
Equipment Failure Analysis
Physiology (medical)
Internal medicine
medicine
Cardiology
Antitachycardia Pacing
Humans
Equipment Failure
Sinus rhythm
R wave amplitude
Cardiology and Cardiovascular Medicine
Intensive care medicine
business
Aged
Subjects
Details
- ISSN :
- 15322092 and 10995129
- Volume :
- 9
- Database :
- OpenAIRE
- Journal :
- EP Europace
- Accession number :
- edsair.doi.dedup.....545d2522aa53fd8d8dd9e7e5903d7b93
- Full Text :
- https://doi.org/10.1093/europace/eum056