Back to Search Start Over

Validation and feasibility of liver T1 mapping using free breathing MOLLI sequence in children and young adults

Authors :
Seung Tae Woo
Seunghyun Lee
Seul Bi Lee
Woo Sun Kim
Young Hun Choi
Jung Eun Cheon
Mun Young Paek
Yeon Jin Cho
Source :
Scientific Reports, Scientific Reports, Vol 10, Iss 1, Pp 1-9 (2020)
Publication Year :
2020
Publisher :
Nature Publishing Group UK, 2020.

Abstract

We investigated the feasibility of free-breathing modified Look-Locker inversion recovery (MOLLI) sequence for measuring hepatic T1 values in children and young adults. To investigate the accuracy and the reproducibility of the T1 maps, a phantom study was performed with 12 different gadoterate meglumine concentrations and the T1 relaxation times of phantoms measured with the MOLLI sequence were compared against those measured with three different sequences: spin-echo inversion recovery, variable flip angle (VFA), and VFA with B1 correction. To evaluate the feasibility of free-breathing MOLLI sequence, hepatic T1 relaxation times obtained by free-breathing and breath-hold technique in twenty patients were compared. The phantom study revealed the excellent accuracy and reproducibility of MOLLI. In twenty patients, the mean value of hepatic T1 values obtained by free-breathing (606.7 ± 64.5 ms) and breath-hold (609.8 ± 64.0 ms) techniques showed no significant difference (p > 0.05). The Bland–Altman plot between the free-breathing and breath-hold revealed that the mean difference of T1 values was − 3.0 ms (− 0.5%). Therefore, T1 relaxation times obtained by MOLLI were comparable to the values obtained using the standard inversion recovery method. The hepatic T1 relaxation times measured by MOLLI technique with free-breathing were comparable to those obtained with breath-hold in children and young adults.

Details

Language :
English
ISSN :
20452322
Volume :
10
Database :
OpenAIRE
Journal :
Scientific Reports
Accession number :
edsair.doi.dedup.....5445328d2f6e10def25b69c471a3ae47