Polyaxial locking and compression screws improve construct stiffness of acetabular cup fixation: a biomechanical study
- Source :
- The Journal of arthroplasty. 29(5)
- Publication Year :
- 2013
-
Abstract
- Bone ingrowth into uncemented acetabular components requires intimate cup-bone contact and rigid fixation, which can be difficult to achieve in revision hip arthroplasty. This study compares polyaxial compression locking screws with non-locked and cancellous screw constructs for acetabular cup fixation. An acetabular cup modified with screw holes to provide both compression and angular stability was implanted into a bone substitute. Coronal lever out, axial torsion and push-out tests were performed with an Instron testing machine, measuring load versus displacement. Polyaxial locking compression screws significantly improved construct stiffness compared with non-locked or cancellous screws. This increased construct stiffness will likely reduce interfacial micromotion. Further research is required to determine whether this will improve bone ingrowth in vivo and reduce cup failure.
- Subjects :
- musculoskeletal diseases
business.product_category
Arthroplasty, Replacement, Hip
Bone Screws
Prosthesis Design
Fixation (surgical)
Osseointegration
medicine
Orthopedics and Sports Medicine
Displacement (orthopedic surgery)
Lever
business.industry
Biomechanics
Stiffness
Acetabulum
musculoskeletal system
equipment and supplies
Compression (physics)
Biomechanical Phenomena
surgical procedures, operative
Coronal plane
Bone Substitutes
Hip Prosthesis
medicine.symptom
business
Biomedical engineering
Subjects
Details
- ISSN :
- 15328406
- Volume :
- 29
- Issue :
- 5
- Database :
- OpenAIRE
- Journal :
- The Journal of arthroplasty
- Accession number :