Visible spectroscopic technique for flowing erythrocytes in capillary

Authors :
Takeshi Shiga
Nobuji Maeda
Norihiko Tateishi
Source :
Biorheology. 27:389-397
Publication Year :
1990
Publisher :
IOS Press, 1990.

Abstract

An optical spectroscopic system for determining the rate of oxygen release from flowing erythrocytes in microvessel is developed. The apparatus consists of following units attached to an inverted microscope. 1) A scanning spectrophotometer, equipped with a grating and a photon counter, was connected to an eyepiece of the microscope through a narrow light-guide, as to obtain the absorption spectrum (wave length range: 450-650 nm) of a focused spot (phi = 7 microns). 2) The velocity of erythrocyte flow was measured by dual-spots cross-correlation method, using two photomultipliers (connected to A/D converter and microcomputer) with two light-guides inserted into another eyepiece. 3) The diameter of vessel was estimated from digitized video-images, using a color image-processor. The ability of the apparatus was tested with (a) hemoglobin solution, (b) flowing erythrocyte suspension and (c) capillaries of rat mesentery. The rate of oxygen release through the vessel wall was calculated.