A simple method for removing initial irregularity of an electrocardiogram during a transient state of a power supply in a defibrillator

Authors :
Jeong-Han Yi
Jin-Su Ahn
Ki-Han Kim
Hyung Sik Kim
Source :
Technology and Health Care
Publication Year :
2020

Abstract

BACKGROUND: The defibrillator is a device that instantaneously discharges the high energy stored in the capacitor to the human body to help revitalize the heart. The circuit for charging the capacitor uses the same power source as the biosignal measurement unit. Therefore, variation in main power supply voltage, ground noise, and electromagnetic interference from the charging circuit can induce distortion into the biosignal at the initial stage of charging. OBJECTIVE: In this study, a simple method is proposed for removing the initial irregularity of an electrocardiogram due to the transient state of a power supply. METHODS: To evaluate the method, a 1-channel electrocardiogram measurement unit and peripheral units were separated from the main control module using galvanic isolation. An isolated push-pull converter was designed to power the secondary side. The method was tested under steady-state and transient conditions. RESULTS: The obtained results proved that biosignal distortion can be significantly reduced. CONCLUSION: This method could be another simple implementation approach for solving signal distortions due to the transient status of power supplies used in medical devices.

Details

ISSN :
18787401
Volume :
28
Issue :
S1
Database :
OpenAIRE
Journal :
Technology and health care : official journal of the European Society for Engineering and Medicine
Accession number :
edsair.doi.dedup.....52fbecd4c932512fc4049711a2f38e89