Minimizing antibody cross-reactivity in multiplex detection of biomarkers in paper-based point-of-care assays

Authors :
Lara Lama
Jesper Gantelius
Helene Andersson-Svahn
Jorge Dias
Source :
Nanoscale. 8(15)
Publication Year :
2016

Abstract

Highly multiplexed immunoassays could allow convenient screening of hundreds or thousands of protein biomarkers simultaneously in a clinical sample such as serum or plasma, potentially allowing improved diagnostic accuracy and clinical management of many conditions such as autoimmune disorders, infections, and several cancers. Currently, antibody microarray-based tests are limited in part due to cross reactivity from detection antibody reagents. Here we present a strategy that reduces the cross-reactivity between nanoparticle-bound reporter antibodies through the application of ultrasound energy. By this concept, it was possible to achieve a sensitivity 10(3)-fold (5 pg mL(-1)) lower than when no ultrasound was applied (50 ng mL(-1)) for the simultaneous detection of three different antigens. The detection limits and variability achieved with this technique rival those obtained with other types of multiplex sandwich assays.

Details

ISSN :
20403372
Volume :
8
Issue :
15
Database :
OpenAIRE
Journal :
Nanoscale
Accession number :