Reproducibility of Anterior Chamber Angle Measurements Obtained with Anterior Segment Optical Coherence Tomography

Authors :
Paul T K Chew
David S. Friedman
Tin Tun Aung
Z. Ce
Yan Li
David Huang
Winifred P. Nolan
Jovina L. See
Scott D. Smith
Sunita Radhakrishnan
Source :
Investigative Opthalmology & Visual Science. 48:3683
Publication Year :
2007
Publisher :
Association for Research in Vision and Ophthalmology (ARVO), 2007.

Abstract

PURPOSE. To evaluate the reproducibility of anterior chamber (AC) angle measurements obtained using anterior segment optical coherence tomography (AS-OCT). METHODS. Patients with suspected glaucoma and those with glaucoma, ocular hypertension, or anatomically narrow angles were recruited from the glaucoma service at the National University Hospital, Singapore. All subjects underwent imaging of the nasal, temporal, and inferior AC angles with an AS-OCT prototype under standardized dark and light conditions. For short-term reproducibility analysis, a single observer acquired two sets of images followed by a third set of images acquired by a second observer. The interval between sessions was 10 minutes. For long-term reproducibility analysis, a single observer acquired two sets of images at least 24 hours apart. Images were measured using custom software to determine the AC depth (ACD), angle opening distance at 500 m (AOD500), angle recess area at 500 m (ARA500), and trabecular–iris space area at 500 m (TISA500). The intraclass correlation coefficient (ICC) was calculated as a measure of intraobserver and interobserver reproducibility. RESULTS. Twenty eyes of 20 patients were analyzed for shortterm reproducibility, and 23 eyes of 23 patients were analyzed for long-term reproducibility. AC depth measurement demonstrated excellent reproducibility (ICC 0.93–1.00) in both dark and light conditions. For the nasal and temporal quadrants, all AC angle parameters demonstrated good to excellent shortterm (ICC 0.67– 0.90) and long-term (ICC 0.56 – 0.93) reproducibility in both dark and light conditions. In the inferior quadrant, reproducibility was lower in all categories of analysis and varied from poor to good (ICC 0.31– 0.73). CONCLUSIONS. AS-OCT allows quantitative assessment of the AC angle. The reproducibility of AC angle measurements was good to excellent for the nasal and temporal quadrants. The lower reproducibility of measurements in the inferior quadrant may be unique to this prototype due to difficulty in acquiring high-quality images of the inferior angle. Further assessment of the commercially available AS-OCT is needed to clarify this finding. (Invest Ophthalmol Vis Sci. 2007;48:3683–3688) DOI: 10.1167/iovs.06-1120

Details

ISSN :
15525783
Volume :
48
Database :
OpenAIRE
Journal :
Investigative Opthalmology & Visual Science
Accession number :
edsair.doi.dedup.....52149e0c615be221784a099c1aa134d6