Novel pedicle screw and plate system provides superior stability in unilateral fixation for minimally invasive transforaminal lumbar interbody fusion: an in vitro biomechanical study
- Source :
- PLoS ONE, Vol 10, Iss 3, p e0123134 (2015), PLoS ONE
- Publication Year :
- 2015
- Publisher :
- Public Library of Science (PLoS), 2015.
-
Abstract
- Purpose This study aims to compare the biomechanical properties of the novel pedicle screw and plate system with the traditional rod system in asymmetrical posterior stabilization for minimally invasive transforaminal lumbar interbody fusion (MI-TLIF). We compared the immediate stabilizing effects of fusion segment and the strain distribution on the vertebral body. Methods Seven fresh calf lumbar spines (L3-L6) were tested. Flexion/extension, lateral bending, and axial rotation were induced by pure moments of ± 5.0 Nm and the range of motion (ROM) was recorded. Strain gauges were instrumented at L4 and L5 vertebral body to record the strain distribution under flexion and lateral bending (LB). After intact kinematic analysis, a right sided TLIF was performed at L4-L5. Then each specimen was tested for the following constructs: unilateral pedicle screw and rod (UR); unilateral pedicle screw and plate (UP); UR and transfacet pedicle screw (TFS); UP and TFS; UP and UR. Results All instrumented constructs significantly reduced ROM in all motion compared with the intact specimen, except the UR construct in axial rotation. Unilateral fixation (UR or UP) reduced ROM less compared with the bilateral fixation (UP/UR+TFS, UP+UR). The plate system resulted in more reduction in ROM compared with the rod system, especially in axial rotation. UP construct provided more stability in axial rotation compared with UR construct. The strain distribution on the left and right side of L4 vertebral body was significantly different from UR and UR+TFS construct under flexion motion. The strain distribution on L4 vertebral body was significantly influenced by different fixation constructs. Conclusions The novel plate could provide sufficient segmental stability in axial rotation. The UR construct exhibits weak stability and asymmetrical strain distribution in fusion segment, while the UP construct is a good alternative choice for unilateral posterior fixation of MI-TLIF.
- Subjects :
- medicine.medical_specialty
Rotation
medicine.medical_treatment
lcsh:Medicine
Lumbar vertebrae
Biomechanical Phenomena
Posterior stabilization
Fixation (surgical)
Pedicle Screws
Lumbar interbody fusion
Bone plate
medicine
Animals
Minimally Invasive Surgical Procedures
Range of Motion, Articular
Pedicle screw
lcsh:Science
Orthodontics
Lumbar Vertebrae
Multidisciplinary
business.industry
lcsh:R
Surgery
Spinal Fusion
medicine.anatomical_structure
Spinal fusion
Cattle
lcsh:Q
business
Bone Plates
Research Article
Subjects
Details
- Language :
- English
- ISSN :
- 19326203
- Volume :
- 10
- Issue :
- 3
- Database :
- OpenAIRE
- Journal :
- PLoS ONE
- Accession number :
- edsair.doi.dedup.....5023ab5fae121cde28765cdedabecfd8