Back to Search
Start Over
High-Resolution Mass Spectrometry Quantification: Impact of Differences in Data Processing of Centroid and Continuum Data
- Source :
- Journal of the American Society for Mass Spectrometry. 30:203-212
- Publication Year :
- 2018
- Publisher :
- American Chemical Society (ACS), 2018.
-
Abstract
- High-resolution mass spectrometry (HRMS) in full scan mode acquires all ions present in the sample of interest offering a lot of qualitative information. This, in combination with the improved performance of the new generation HRMS systems, triggers more (bio) analysts to switch from triple quad MS systems to HRMS for quantitative analysis. Quantitative processing of HRMS data is performed based on narrow mass extraction windows rather than on nominal mass product ion chromatograms (SRM or MRM). Optimal processing of HRMS data requires different considerations and software tools and can have an impact on data processing and final results. The selection of centroid versus continuum/profile data for processing, selection of the optimal narrow mass extraction window, using theoretical versus measured accurate mass for the extraction of the ion chromatograms as well as differences in calculations and data handling residing in the different vendor software packages are tackled in the presented manuscript. These differences are illustrated on HRMS data acquired for the same plasma samples on three different platforms, i.e., a Sciex QToF, a Waters QToF, and a Thermo Orbitrap system, and processed in four different software packages, i.e., Sciex Analyst® TF, Waters Masslynx, Waters Unifi, and Thermo Xcalibur. The impact of these differences on quantitative HRMS performance was evaluated on calibration curves of eight small molecule compounds in plasma using four different ways of processing. Simple guidelines are provided for the selection of an optimal mass extraction window for continuum and centroided data. Graphical Abstract.
- Subjects :
- Data processing
Group method of data handling
Chemistry
business.industry
Calibration curve
010401 analytical chemistry
Centroid
010402 general chemistry
Orbitrap
Mass spectrometry
01 natural sciences
0104 chemical sciences
law.invention
Mass
Software
Structural Biology
law
business
Biological system
Spectroscopy
Subjects
Details
- ISSN :
- 10440305
- Volume :
- 30
- Database :
- OpenAIRE
- Journal :
- Journal of the American Society for Mass Spectrometry
- Accession number :
- edsair.doi.dedup.....4f1e649bed7fa86d8efbc6857763d795
- Full Text :
- https://doi.org/10.1007/s13361-018-2101-0