In Vitro and In Vivo Testing of an Implantable Motor-Driven Left Ventricular Assist Device

Authors :
Yuichiro Matsuura
Tatsuki Katsumura
Shintaro Kukunaga
Takashi Fujiwara
Daiki Kikugawa
Yoshiaki Fukuhiro
Shigeo Kanazawa
Taiji Murakami
Source :
Artificial Organs. 20:152-155
Publication Year :
2008
Publisher :
Wiley, 2008.

Abstract

A totally implantable motor-driven left ventricular assist device (LVAD) has been developed and tested. The performance of this LVAD was tested in a mock circulatory system. This pump provided 8 L/min of output against a mean afterload of 120 mm Hg with a filling pressure of 20 mm Hg when the pump was operated in the fill/empty mode. The right and left pumps were tested in a mock loop. The right pump afterload was kept in the range from 23–32 mm Hg. With increase in the left pump afterload, the pump power output varied from 1.64 to 2.37 W. The instantaneous motor power input varied from 22.6 to 30.6 W with the total system efficiency ranging from 6.7 to 9.4%. To date, 4 in vivo studies have been conducted for up to 12 h. Two animals survived 12 and 10 h, respectively. Termination was due to bleeding in 1 animal, vent tube obstruction in 1, and respiratory failure in 2. All animals died of technical failure. Another experiment is to be undertaken, and a newly designed cannula is now being manufactured.

Details

ISSN :
15251594 and 0160564X
Volume :
20
Database :
OpenAIRE
Journal :
Artificial Organs
Accession number :
edsair.doi.dedup.....4e2c38d5791bd7c3c12c4226917e5373
Full Text :