Frequent Suboptimal Thermocycler Ramp Rate Usage Negatively Impacts GenoType MTBDRsl VER 2.0 Performance for Second-Line Drug-Resistant Tuberculosis Diagnosis

Authors :
Brigitta Derendinger
Margaretha de Vos
Samantha Pillay
Rouxjeane Venter
John Metcalfe
Yonas Ghebrekristos
Stephanie Minnies
Tania Dolby
Natalie Beylis
Robin Warren
Grant Theron
Source :
J Mol Diagn
Publication Year :
2022
Publisher :
Elsevier BV, 2022.

Abstract

Strengthening second-line drug-resistant tuberculosis (TB) detection is a priority. GenoType MTBDRplus VER 2.0 performance is reduced with non-recommended ramp rate usage (temperature change speed between PCR cycles); however, ramp rate’s effect on GenoType MTBDRsl VER 2.0 (MTBDRsl) performance, is unknown. Fifty-two Xpert MTB/RIF Ultra-positive rifampicin-resistant smear-negative sputa and a Mycobacterium tuberculosis dilution series were tested at a manufacturer-recommended (2.2°C/second) or suboptimal (4.0°C/second) ramp rate. M. tuberculosis–complex-DNA positivity, indeterminates, fluoroquinolone- and second-line injectable-resistance accuracy, banding differences, and, separately, inter-reader variability were assessed. Five (39%) of 13 re-surveyed laboratories did not use the manufacturer-recommended ramp rate. On sputum, 2.2°C/second improved indeterminates versus 4.0°C/second (0 of 52 versus 7 of 51; P = 0.006), incorrect drug-class diagnostic calls (0 of 104 versus 6 of 102; P = 0.013), and incorrect banding calls (0 of 1300 versus 54 of 1275; P < 0.001). Similarly, 2.2°C/second improved valid results [(52 of 52 versus 41 of 51; +21% (P = 0.001)] and banding call inter-reader variability [34 of 1300 (3%) versus 52 of 1300 (4%); P = 0.030]. At the suboptimal ramp rate, false-resistance and false-susceptible calls resulted from wild-type band absence rather than mutant band appearance, resulting in misclassification of moxifloxacin resistance level from high-to-low. Suboptimal ramp rate contributes to poor MTBDRsl performance. Laboratories must ensure that the manufacturer-recommended ramp rate is used.

Details

ISSN :
15251578
Volume :
24
Database :
OpenAIRE
Journal :
The Journal of Molecular Diagnostics
Accession number :
edsair.doi.dedup.....4e01dc4f7e601d17fff5cf77e66d7001
Full Text :
https://doi.org/10.1016/j.jmoldx.2022.01.003