Accuracy in positioning of dental X-ray images – A comparative study of a portable X-ray device and a wall-mounted device

Authors :
Lara Schorn
Uwe Zeller
Julian Lommen
Julia Nitschke
Christoph Sproll
Jörg Handschel
Norbert R. Kübler
Henrik Holtmann
Source :
Saudi Dental Journal, Vol 33, Iss 8, Pp 884-889 (2021), The Saudi Dental Journal
Publication Year :
2021
Publisher :
Elsevier, 2021.

Abstract

Introduction: The benefits of portable dental X-ray devices remain controversially debated. This study aimed to compare the accuracy in positioning dental X-ray images using handheld (Nomad Pro 2) and wall-mounted (Heliodent Plus) X-ray devices. Materials and methods: Radiographical imaging was exercised on a maxillofacial phantom using the handheld dental X-ray device Nomad Pro 2 (Kavo Kerr, Biberach, Germany) and the wall-mounted dental X-ray device Heliodent Plus (Sirona Dental Systems, Bensheim, Germany). Accuracy of device positioning (i.e., centeredness and perpendicularity) was measured as horizontal and vertical deviation (pixels and millimeters) from a centrally positioned crosshair. The reproducibility of the results was tested for 80 images per device and operator. IBM SPSS (SPSS Inc., Chicago, IL) was used for statistical analyses. Results: Centeredness of produced dental X-rays was significantly better for the wall-mounted device than for the handheld device in both deviation from the x-axis (p = 0.042) and y-axis (p = 0.020). The perpendicularity of the produced dental X-rays was significantly better for the handheld device than for the wall-mounted device for both horizontal (p

Details

Language :
English
ISSN :
10139052
Volume :
33
Issue :
8
Database :
OpenAIRE
Journal :
Saudi Dental Journal
Accession number :
edsair.doi.dedup.....4de311719d897e7792cce65e48da331b