Accuracy in positioning of dental X-ray images – A comparative study of a portable X-ray device and a wall-mounted device
- Source :
- Saudi Dental Journal, Vol 33, Iss 8, Pp 884-889 (2021), The Saudi Dental Journal
- Publication Year :
- 2021
- Publisher :
- Elsevier, 2021.
-
Abstract
- Introduction: The benefits of portable dental X-ray devices remain controversially debated. This study aimed to compare the accuracy in positioning dental X-ray images using handheld (Nomad Pro 2) and wall-mounted (Heliodent Plus) X-ray devices. Materials and methods: Radiographical imaging was exercised on a maxillofacial phantom using the handheld dental X-ray device Nomad Pro 2 (Kavo Kerr, Biberach, Germany) and the wall-mounted dental X-ray device Heliodent Plus (Sirona Dental Systems, Bensheim, Germany). Accuracy of device positioning (i.e., centeredness and perpendicularity) was measured as horizontal and vertical deviation (pixels and millimeters) from a centrally positioned crosshair. The reproducibility of the results was tested for 80 images per device and operator. IBM SPSS (SPSS Inc., Chicago, IL) was used for statistical analyses. Results: Centeredness of produced dental X-rays was significantly better for the wall-mounted device than for the handheld device in both deviation from the x-axis (p = 0.042) and y-axis (p = 0.020). The perpendicularity of the produced dental X-rays was significantly better for the handheld device than for the wall-mounted device for both horizontal (p
Details
- Language :
- English
- ISSN :
- 10139052
- Volume :
- 33
- Issue :
- 8
- Database :
- OpenAIRE
- Journal :
- Saudi Dental Journal
- Accession number :
- edsair.doi.dedup.....4de311719d897e7792cce65e48da331b