Simultaneous pre-concentration and separation on simple paper-based analytical device for protein analysis
- Source :
- Analytical and Bioanalytical Chemistry. 410:1689-1695
- Publication Year :
- 2018
- Publisher :
- Springer Science and Business Media LLC, 2018.
-
Abstract
- In this work, fast isoelectric focusing (IEF) was successfully implemented on an open paper fluidic channel for simultaneous concentration and separation of proteins from complex matrix. With this simple device, IEF can be finished in 10 min with a resolution of 0.03 pH units and concentration factor of 10, as estimated by color model proteins by smartphone-based colorimetric detection. Fast detection of albumin from human serum and glycated hemoglobin (HBA1c) from blood cell was demonstrated. In addition, off-line identification of the model proteins from the IEF fractions with matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF-MS) was also shown. This PAD IEF is potentially useful either for point of care test (POCT) or biomarker analysis as a cost-effective sample pretreatment method.
- Subjects :
- Glycated Hemoglobin
Paper
Time Factors
Chromatography
Resolution (mass spectrometry)
Chemistry
Isoelectric focusing
010401 analytical chemistry
Albumin
Serum Albumin, Human
Equipment Design
Paper based
010402 general chemistry
Mass spectrometry
01 natural sciences
Biochemistry
0104 chemical sciences
Analytical Chemistry
Point-of-Care Testing
Lab-On-A-Chip Devices
Humans
Concentration factor
Biomarker Analysis
Isoelectric Focusing
Pre concentration
Subjects
Details
- ISSN :
- 16182650 and 16182642
- Volume :
- 410
- Database :
- OpenAIRE
- Journal :
- Analytical and Bioanalytical Chemistry
- Accession number :
- edsair.doi.dedup.....4d0ea6cf3cd5a68ae19d0fa73620113e
- Full Text :
- https://doi.org/10.1007/s00216-017-0809-5