Back to Search Start Over

Optical and Electrical Method Characterizing the Dynamic Behavior of the Fused Silica Cylindrical Resonator

Authors :
Tianliang Qu
Jie Yuan
Yonglei Jia
Kaiyong Yang
Zhinan Qiu
Yao Pan
Zhenfang Fan
Hui Luo
Source :
Sensors (Basel, Switzerland), Sensors, Vol 19, Iss 13, p 2928 (2019), Sensors, Volume 19, Issue 13
Publication Year :
2019
Publisher :
MDPI AG, 2019.

Abstract

Fused silica cylindrical resonant gyroscope (CRG) is a novel high-precision solid-wave gyroscope, whose performance is primarily determined by the cylindrical resonator&rsquo<br />s frequency split and quality factor (Q factor). The laser Doppler vibrometer (LDV) is extensively used to measure the dynamic behavior of fused silica cylindrical resonators. An electrical method was proposed to characterize the dynamic behavior of the cylindrical resonator to enhance the measurement efficiency and decrease the equipment cost. With the data acquisition system and the designed signal analysis program based on LabVIEW software, the dynamic behavior of the fused silica cylindrical resonator can be analyzed automatically and quickly. We compared all the electrical measurement results with the optical detection by LDV, demonstrating that the fast Fourier transform (FFT) result of the resonant frequency measured by the electrical method was 0.12 Hz higher than that with the optical method. Thus, the frequency split measured by the electrical and optical methods was the same in 0.18 Hz, and the measurement of the Q factor was basically the same in 730,000. We conducted all measurements under the same operation condition, and the optical method was used as a reference, demonstrating that the electrical method could characterize the dynamic behavior of the fused silica cylindrical resonator and enhance the measurement efficiency.

Details

ISSN :
14248220
Volume :
19
Database :
OpenAIRE
Journal :
Sensors
Accession number :
edsair.doi.dedup.....4c9ac5e1faa2eedd3d949ec2a7248f51