Back to Search Start Over

The hard X-ray Photon Single-Shot Spectrometer of SwissFEL—initial characterization

Authors :
Carla David
P. Wiegand
P. Heimgartner
Mikako Makita
Pavle Juranić
L. Patthey
Christof W. Schneider
U. Flechsig
V. Thominet
A. Rodriguez Fernandez
G. Seniutinas
C. Pradervand
J. Rehanek
Source :
Journal of Instrumentation
Publication Year :
2017
Publisher :
IOP Publishing, 2017.

Abstract

SwissFEL requires the monitoring of the photon spectral distribution at a repetition rate of 100 Hz for machine optimization and experiment online diagnostics. The Photon Single Shot Spectrometer has been designed for the photon energy range of 4 keV to 12 keV provided by the Aramis beamline. It is capable of measuring the spectrum in a non-destructive manner, with an energy resolution of delta(E)/E = (2-5) x 10-5 over a bandwidth of 0.5% on a shot-to-shot basis. This article gives a detailed description about the technical challenges, structures, and considerations when building such a device, and to further enhance the performance of the spectrometer.

Details

ISSN :
17480221 and 17426596
Volume :
12
Database :
OpenAIRE
Journal :
Journal of Instrumentation
Accession number :
edsair.doi.dedup.....4bc46f824820b602f54cbb9c3bc25f86
Full Text :
https://doi.org/10.1088/1748-0221/12/05/p05024