Back to Search Start Over

Spatial resolution of a spherical x-ray crystal spectrometer at various magnifications

Authors :
P. C. Efthimion
E. O. Baronova
K. W. Hill
L. F. Delgado-Aparicio
Manfred Bitter
N. A. Pablant
Lan Gao
N. R. Pereira
Source :
The Review of scientific instruments. 87(11)
Publication Year :
2016

Abstract

A high spatial resolution of a few μm is often required for probing small-scale high-energy-density plasmas using high resolution x-ray imaging spectroscopy. This resolution can be achieved by adjusting system magnification to overcome the inherent limitation of the detector pixel size. Laboratory experiments on investigating the relation between spatial resolution and system magnification for a spherical crystal spectrometer are presented. Tungsten Lβ2 rays from a tungsten-target micro-focus x-ray tube were diffracted by a Ge 440 crystal, which was spherically bent to a radius of 223 mm, and imaged onto an x-ray CCD with 13-μm pixel size. The source-to-crystal (p) and crystal-to-detector (q) distances were varied to produce spatial magnifications (M = q/p) ranging from 2 to 10. The inferred instrumental spatial width reduces with increasing system magnification M. However, the experimental measurement at each M is larger than the theoretical value of pixel size divided by M. Future work will focus on investigating possible broadening mechanisms that limit the spatial resolution.

Details

ISSN :
10897623
Volume :
87
Issue :
11
Database :
OpenAIRE
Journal :
The Review of scientific instruments
Accession number :