Isotopic Analysis by Laser Ablation Solution Sampling MC-ICP-MS─An Example of Boron

Authors :
Xiuhong Liao
Zhaochu Hu
Wen Zhang
Yantong Feng
Tao Luo
Zaicong Wang
Ming Li
Keqing Zong
Yongsheng Liu
Shenghong Hu
Source :
Analytical Chemistry. 94:1286-1293
Publication Year :
2021
Publisher :
American Chemical Society (ACS), 2021.

Abstract

Using boron as a test analyte, laser ablation (LA) solution sampling multicollector inductively coupled plasma mass spectrometry (MC-ICP-MS) is proposed and validated as a fast method for isotopic analysis in natural liquids and digested samples without any prior purification process. We demonstrated that the solution reference standard can be used as a bracketing standard for in situ δ