Reporting Insulin Pump Accuracy: Trumpet Curves According to IEC 60601-2-24 and Beyond
- Source :
- Journal of Diabetes Science and Technology. 13:592-596
- Publication Year :
- 2018
- Publisher :
- SAGE Publications, 2018.
-
Abstract
- Accuracy of insulin pump basal rate delivery, if tested according to the standard IEC 60601-2-24 for infusion pumps, shall be presented as a trumpet curve. This way of graphical presentation is common; however, it is often misunderstood and misinterpreted by people. It is often assumed that a trumpet curve shows the error rate as a function of time, thus implying an increasing accuracy in the course of time. On the contrary, the horizontal axis of a trumpet curve shows increasingly long observation windows. In addition, trumpet curves display only extreme values, that is, those windows with minimal and maximal deviation, which might not be representative for the total deviation. This commentary provides information regarding the calculations and the interpretation of trumpet curves and proposes alternative approaches.
- Subjects :
- Insulin pump
medicine.medical_specialty
Basal rate
Endocrinology, Diabetes and Metabolism
Biomedical Engineering
030209 endocrinology & metabolism
Bioengineering
03 medical and health sciences
Insulin Infusion Systems
0302 clinical medicine
Commentaries
Germany
Internal medicine
Computer Graphics
Diabetes Mellitus
Internal Medicine
medicine
Humans
Insulin
Drug Dosage Calculations
030212 general & internal medicine
Medical Device Legislation
Infusion Pumps
business.industry
Reproducibility of Results
Reference Standards
Data Accuracy
Equipment and Supplies
Research Design
Data Interpretation, Statistical
Cardiology
Presentation (obstetrics)
business
Subjects
Details
- ISSN :
- 19322968
- Volume :
- 13
- Database :
- OpenAIRE
- Journal :
- Journal of Diabetes Science and Technology
- Accession number :
- edsair.doi.dedup.....474e369360267b191114bd876fcab98f
- Full Text :
- https://doi.org/10.1177/1932296818806509