In vivo crystalline lens measurements with novel swept-source optical coherent tomography: an investigation on variability of measurement

Authors :
Norihiro Yamada
Kei Shinoda
Itaru Kimura
Takuhei Shoji
Sho Ishikawa
Naoko Kato
Hisashi Ibuki
Source :
BMJ Open Ophthalmology
Publication Year :
2017
Publisher :
BMJ Open Ophthalmology, 2017.

Abstract

Objective To evaluate the reproducibility of in vivo crystalline lens measurements obtained by novel commercially available swept-source (SS) optical coherence tomography (OCT) specifically designed for anterior segment imaging. Methods and analysis One eye from each of 30 healthy subjects was randomly selected using the CASIA2 (Tomey, Nagoya, Japan) in two separate visits within a week. Each eye was imaged twice. After image scanning, the anterior and posterior lens curvatures and lens thickness were calculated automatically by the CASIA2 built-in program at 0 dioptre (D) (static), −1 D, −3 D and −5 D accommodative stress. The intraobserver and intervisit reproducibility coefficient (RC) and intraclass correlation coefficient (ICC) were calculated. Results The intraobserver and intervisit RCs ranged from 0.824 to 1.254 mm and 0.789 to 0.911 mm for anterior lens curvature, from 0.276 to 0.299 mm and 0.221 to 0.270 mm for posterior lens curvature and from 0.065 to 0.094 mm and 0.054 to 0.132 mm for lens thickness, respectively. The intraobserver and intervisit ICCs ranged from 0.831 to 0.865 and 0.828 to 0.914 for anterior lens curvature, from 0.832 to 0.898 and 0.840 to 0.933 for posterior lens curvature and from 0.980 to 0.992 and 0.942 to 0.995 for lens thickness. High ICC values were observed for each measurement regardless of accommodative stress. RCs in younger subjects tended to be larger than those in older subjects. Conclusions This novel anterior segment SS-OCT instrument produced reliable in vivo crystalline lens measurement with good repeatability and reproducibility regardless of accommodation stress.

Details

Language :
English
ISSN :
23973269
Volume :
1
Issue :
1
Database :
OpenAIRE
Journal :
BMJ Open Ophthalmology
Accession number :
edsair.doi.dedup.....46255aa1f9731b6b4b2550430348c3e4