Back to Search
Start Over
Multicenter Reproducibility of Liver Iron Quantification with 1.5-T and 3.0-T MRI
- Source :
- Radiology.
- Publication Year :
- 2022
-
Abstract
- Background MRI is a standard of care tool to measure liver iron concentration (LIC). Compared with regulatory-approved R2 MRI, R2* MRI has superior speed and is available in most MRI scanners; however, the cross-vendor reproducibility of R2*-based LIC estimation remains unknown. Purpose To evaluate the reproducibility of LIC via single-breath-hold R2* MRI at both 1.5 T and 3.0 T with use of a multicenter, multivendor study. Materials and Methods Four academic medical centers using MRI scanners from three different vendors (three 1.5-T scanners, one 2.89-T scanner, and two 3.0-T scanners) participated in this prospective cross-sectional study. Participants with known or suspected liver iron overload were recruited to undergo multiecho gradient-echo MRI for R2* mapping at 1.5 T and 3.0 T (2.89 T or 3.0 T) on the same day. R2* maps were reconstructed from the multiecho images and analyzed at a single center. Reference LIC measurements were obtained with a commercial R2 MRI method performed using standardized 1.5-T spin-echo imaging. R2*-versus-LIC calibrations were generated across centers and field strengths using linear regression and compared using
- Subjects :
- Radiology, Nuclear Medicine and imaging
Subjects
Details
- ISSN :
- 15271315
- Database :
- OpenAIRE
- Journal :
- Radiology
- Accession number :
- edsair.doi.dedup.....441e98b38e30cfa10757261bdaa450cb