Wear of a 5 Megarad Cross-linked Polyethylene Liner: A 6-year RSA Study
- Source :
- Clinical Orthopaedics & Related Research. 471:2238-2244
- Publication Year :
- 2013
- Publisher :
- Ovid Technologies (Wolters Kluwer Health), 2013.
-
Abstract
- One cross-linked polyethylene (XLPE) liner is manufactured using a lower dose of radiation, 5 Mrad, which may result in less cross-linking. The reported in vivo wear rate of this XLPE liner in patients undergoing THA has varied, and has included some patients in each reported cohort who had greater than 0.1 mm/year of wear, which is an historical threshold for osteolysis. Previous studies have measured wear on plain radiographs, an approach that has limited sensitivity.We therefore measured the amount and direction of wear at 6 years using Radiostereometric analysis (RSA) in patients who had THAs that included a cross-linked polyethylene liner manufactured using 5 Mrad radiation.We prospectively reviewed wear in 30 patients who underwent primary THAs with the same design of cross-linked acetabular liner and a 28-mm articulation. Tantalum markers were inserted during surgery and all patients had RSA radiographic examinations at 1 week, 6 months, 1, 2, and 6 years postoperatively.The mean proximal, two-dimensional (2-D) and three-dimensional (3-D) wear rates calculated between 1 year and 6 years were 0.014, 0.014, and 0.018 mm/per year, respectively. The direction of the head penetration recorded between 1 week and 6 years was in a proximal direction for all patients, proximolateral for 16 of 24 patients, and proximomedial for eight of 24 patients.The proximal, 2-D and 3-D wear of a XLPE liner produced using 5 Mrad of radiation was low but measurable by RSA after 6 years. No patients had proximal 2-D or 3-D wear rates exceeding 0.1 mm/year. Further followup is needed to evaluate the effect of XLPE wear particles on the development of long-term osteolysis.
- Subjects :
- Male
musculoskeletal diseases
medicine.medical_specialty
Time Factors
Arthroplasty, Replacement, Hip
Treatment outcome
Osteolysis
Tantalum
Prosthesis Design
Radiostereometric Analysis
chemistry.chemical_compound
Clinical Research
Fiducial Markers
medicine
Humans
Prosthesis design
Orthopedics and Sports Medicine
Prospective Studies
Composite material
Aged
Aged, 80 and over
Cross-linked polyethylene
business.industry
technology, industry, and agriculture
Polyethylene liner
General Medicine
Middle Aged
Polyethylene
equipment and supplies
Prosthesis Failure
Surgery
Treatment Outcome
chemistry
Female
Hip Joint
Hip Prosthesis
Stress, Mechanical
business
Subjects
Details
- ISSN :
- 0009921X
- Volume :
- 471
- Database :
- OpenAIRE
- Journal :
- Clinical Orthopaedics & Related Research
- Accession number :
- edsair.doi.dedup.....43e29bf4fe0d097a2d5335c2c0b64eb9
- Full Text :
- https://doi.org/10.1007/s11999-013-2789-x