Back to Search Start Over

Wear of a 5 Megarad Cross-linked Polyethylene Liner: A 6-year RSA Study

Authors :
David G. Campbell
Stuart A. Callary
Graham Mercer
Kjell G. Nilsson
John R. Field
Source :
Clinical Orthopaedics & Related Research. 471:2238-2244
Publication Year :
2013
Publisher :
Ovid Technologies (Wolters Kluwer Health), 2013.

Abstract

One cross-linked polyethylene (XLPE) liner is manufactured using a lower dose of radiation, 5 Mrad, which may result in less cross-linking. The reported in vivo wear rate of this XLPE liner in patients undergoing THA has varied, and has included some patients in each reported cohort who had greater than 0.1 mm/year of wear, which is an historical threshold for osteolysis. Previous studies have measured wear on plain radiographs, an approach that has limited sensitivity.We therefore measured the amount and direction of wear at 6 years using Radiostereometric analysis (RSA) in patients who had THAs that included a cross-linked polyethylene liner manufactured using 5 Mrad radiation.We prospectively reviewed wear in 30 patients who underwent primary THAs with the same design of cross-linked acetabular liner and a 28-mm articulation. Tantalum markers were inserted during surgery and all patients had RSA radiographic examinations at 1 week, 6 months, 1, 2, and 6 years postoperatively.The mean proximal, two-dimensional (2-D) and three-dimensional (3-D) wear rates calculated between 1 year and 6 years were 0.014, 0.014, and 0.018 mm/per year, respectively. The direction of the head penetration recorded between 1 week and 6 years was in a proximal direction for all patients, proximolateral for 16 of 24 patients, and proximomedial for eight of 24 patients.The proximal, 2-D and 3-D wear of a XLPE liner produced using 5 Mrad of radiation was low but measurable by RSA after 6 years. No patients had proximal 2-D or 3-D wear rates exceeding 0.1 mm/year. Further followup is needed to evaluate the effect of XLPE wear particles on the development of long-term osteolysis.

Details

ISSN :
0009921X
Volume :
471
Database :
OpenAIRE
Journal :
Clinical Orthopaedics & Related Research
Accession number :
edsair.doi.dedup.....43e29bf4fe0d097a2d5335c2c0b64eb9
Full Text :
https://doi.org/10.1007/s11999-013-2789-x